Breakaway Forces of Flat and Domed Surfaced Magfittm Implant Magnet Attachments

Abstract - The aim of this in vitro study was to compare the resistance to separation (breakaway force) of flat (Magfit™-IP-BF) and domed (Magfit™-IP-BD) implant magnet attachments subjected to non-axial forces. The separating forces were applied by an Instron Universal Testing Instrument to single magnet attachments at angles of 0°, 2°, 5°, 10°, and 20° from the axial line of the components (angle of pull) and at crosshead speeds of 0.5mm/min and 50 mm/min. The breakaway forces were significantly (p<0.0001) inversely related to the angle of pull for both flat magnets and for domed magnets. At the slow crosshead speed, the breakaway forces recorded for the domed magnets were significantly greater than those recorded for the flat magnets for angles of pull greater than 5°. At the faster crosshead speed, the breakaway forces recorded for the domed magnets were significantly greater than those recorded for the flat magnets for angles of pull greater than 2°. This apparent superiority of domed magnets under non-axially directed separating forces could influence the choice of magnet attachment for implant overdentures as intraoral displacing forces are multidirectional. Domed magnets may also be advantageous where implants are not parallel.

KEY WORDS: Breakaway force, angle of pull, implant magnet attachments, flat magnets, dome magnets, overdentures

Article Price
£15.00
Institution Article Price
£
Page Start
7
Page End
12
Authors
V. Chopra, B. J. Smith, H. W. Preiskel, R. M. Palmer, R. Curtis

Articles from this issue

  • Title
  • Pg. Start
  • Pg. End

  1. Long-term Survival of Direct and Indirect Restorations Placed for the Treatment of Advanced Tooth Wear
  2. 2
  3. 6

  1. Breakaway Forces of Flat and Domed Surfaced Magfittm Implant Magnet Attachments
  2. 7
  3. 12

  1. The Abrasion of Dental Composite by CobaltChromium Clasps
  2. 13
  3. 18

  1. Fracture Resistance of Roots with Thin Walls Restored Using an Intermediate Resin Composite Layer Placed Between the Dentine and a Cast Metal Post
  2. 19
  3. 22

  1. The Effect of a Range of Disinfectants on the Dimensional Accuracy and Stability of Some Impression Materials
  2. 23
  3. 28

  1. Mathematical Analysis of Occlusal Rest Design for Cast Removable Partial Dentures
  2. 29
  3. 32

  1. Analytical and Electrochemical Evaluation of the in vitro corrosion behavior of Nickel-chrome and Cobaltchrome casting alloys for metal-ceramic restorations.
  2. 33
  3. 40

  1. Apexification with Mineral Trioxide Aggregate: an overview of the material and technique
  2. 41
  3. 45